Remove Chiller Vents and Thumb Nuts
- Remove both chiller vent panels.
- Unscrew all 5 thumb nuts:
- 3 from the left chiller vent
- 2 from the right chiller vent
-
Note: Bottom-right nut may be an 8mm hex nut—use socket if necessary.
|
|
Remove Rear Cover
- Slowly pull and wiggle the rear cover out through the vent openings.
|
|
Remove E-Box Cover
- Use a 2.5mm hex screwdriver to remove the 4 screws securing the e-box cover.
|
|
-
Unplug & Remove Flow Meter From Ebox
-
Using the PCB Board Diagram, locate and unplug the flow meter from the PCB board.
-
NOTE: The flow meter plug is too large to fit through the port with other wires in place, so you'll need to first remove the TH1 thermistor and suction thermistor.
- Unplug the TH1 and suction thermistors from the PCB board.
- Push the TH1 and suction thermistor connectors through the back port of the electrical box.
-
NOTE: Do not pull on the wires. Only push the plastic connectors through to avoid damaging the wires or connections.
- Once those are removed, push the flow meter plug out through the back of the electrical box.
- Leave all three connectors outside the electrical box in preparation for installing the new flow meter manifold.
|
|
Cut Zip Ties Securing Flow Meter Wire
- Cut the zip ties securing the flow meter wire to the chiller.
|
|
Disconnect Flow Meter Manifold From Water Filter Housing
- Locate the ab union connecting the flow meter manifold to the back of the water filter housing.
- Unscrew the ab union to disconnect the flow meter manifold from the water filter housing
NOTE: If a hand tool is necessary to unscrew ab union you must use soft tip pliers. |
|
Remove Flow Meter From Manifold
- Locate the flow meter on the flow meter manifold between the water filter housing and fan.
- Using a phillips head screwdriver remove the two screws securing the cover of the flow meter.
- Using a 3mm hex bolt driver remove the 4 bolts securing the flow meter to the manifold.
- Cut zip ties that are securing the flow meter the the chiller
|
|
Install New Flow Meter to Manifold
- Before installing the flow meter ensure the pin wheel on the back side spins freely without obstruction.
-
NOTE: There is a plastic pin that holds the pin wheel in place. Be careful this does not fall out.
- Using a 3mm hex nut driver install the flow meter into the flow meter manifold.
-
NOTE: The wire should be facing up to the top of the tub when the flow meter is installed
- Using a Phillips head screw driver install and secure the flow meter cover to the flow meter.
|
|
Install & Plug In Flow Meter to Ebox
- Feed the flow meter, water temp thermistor, and suction thermistor wires through the back of the electrical box.
-
NOTE: The flow meter must go first due to the size of the connector.
-
Refer to the PCB board diagram and plug each wire into the correct port:
⚠️ Important: Plugging any of these into the wrong port can cause the chiller to malfunction or become damaged. Always double-check against the PCB diagram before powering the unit back on. |
|
Secure Flow Meter Cord Using Zip Ties
- Secure the flow meter wire to the other cords along the top of the fan.
|
|
Connect Flow Meter Manifold From Water Filter Housing
- Connect the flow meter union to the back of the water filter housing connection.
⚠️ Important: Hand tighten only, tools will damage the union. If the union is difficult to screw on back out as it can be cross threaded.
|
|
Test for Leaks & CF Error Code
⚠️ Warning: The chiller fan is exposed during testing. Keep hands, tools, and loose items away from the fan to avoid injury or equipment damage.
- Fill the Plunge up with water so the water line is 6" below the edge of the tub.
-
🚨 Open the auto-doser ball valves🚨
- ⚠️ If chiller is turned on while auto-doser ball valves are closed it can damage the chiller.⚠️
- Plug in the Plunge and press the “Reset” button on the GFCI plug to power it on.
- Set temperature to 50 degrees so pump powers to 100%.
- Allow the Plunge to run and observe the chiller plumbing union connections.
- Carefully inspect all union fittings for any signs of leaking while the system is running.
- If a leak is found, pause the system and recheck that the union is properly seated and hand-tightened.
- Confirm no "CF" error code on UI
|
|
Power Down The Plunge
- Pause the chiller by pressing the pause button on the UI
- ⚠️ Powering down the Plunge while it's in a cooling cycle can damage the chiller.
- Press the test button on the GFCI outlet to power down the Plunge.
- Unplug the GFCI outlet from the outlet.
🚨 Important: Do not power down the Plunge before pausing the chiller.
|
|
Prepare for Rear Cover Installation
- Reattach e-box cover (4 screws).
- Secure chiller to frame using the 4 base nuts.
- Pre-install the upper thumb nut (near filter housing) loosely—this area is tight with cover on.
|
|
Install Rear Cover
- Carefully slide rear cover over chiller.
- Ensure power cord is routed through the correct side.
- Reinstall all 5 thumb nuts to secure rear panel.
- Attach the left and right chiller vents.
|
|